The multifocal visual evoked cortical potential in visual field mapping: a methodological study.

Macfarlane, Jennifer A (2008) The multifocal visual evoked cortical potential in visual field mapping: a methodological study. PhD thesis, University of Glasgow.

Full text available as:
[thumbnail of 2008macfarlanephd.pdf] PDF
Download (2MB)
Printed Thesis Information: https://eleanor.lib.gla.ac.uk/record=b2636837

Abstract

The application of multifocal techniques to the visual evoked cortical potential permits objective electrophysiological mapping of the visual field.
The multifocal visual evoked cortical potential (mfVECP) presents several technical challenges. Signals are small, are influenced by a number of sources of noise and waveforms vary both across the visual field and between subjects due to the complex geometry of the visual cortex. Together these factors hamper the ability to distinguish between a mfVECP response from the healthy visual pathway, and a response that is reduced or absent and is therefore representative of pathology.
This thesis presents a series of methodological investigations with the aim of maximising the information available in the recorded electrophysiological response, thereby improving the performance of the mfVECP.
A novel method of calculating the signal to noise ratio (SNR) of mfVECP waveform responses is introduced. A noise estimate unrelated to the response of the visual cortex to the visual stimulus is created. This is achieved by cross-correlating m-sequences which are created when the orthogonal set of m-sequences are created but are not used to control a stimulus region, with the physiological record. This metric is compared to the approach of defining noise within a delayed time window and shows good correlation. ROC analysis indicates a small improvement in the ability to distinguish between physiological waveform responses and noise. Defining the signal window as 45-250ms is recommended.
Signal quality is improved by post-acquisition bandwidth filtering. A wide range of bandwidths are compared and the greatest gains are seen with a bandpass of 3 to 20Hz applied after cross-correlation.
Responses evoked when stimulation is delivered using a cathode ray tube (CRT) and a liquid crystal display (LCD) projector system are compared. The mode of stimulus delivery affects the waveshape of responses. A significantly higher SNR is seen in waveforms is shown in waveforms evoked by an m=16 bit m-sequence delivered by a CRT monitor. Differences for shorter m-sequences were not statistically significant.
The area of the visual field which can usefully be tested is investigated by increasing the field of view of stimulation from 20° to 40° of radius in 10° increments. A field of view of 30° of radius is shown to provide stimulation of as much of the visual field as possible without losing signal quality.
Stimulation rates of 12.5 to 75Hz are compared. Slowing the stimulation rate produced increases waveform amplitudes, latencies and SNR values. The best performance was achieved with 25Hz stimulation. It is shown that a six-minute recording stimulated at 25Hz is superior to an eight-minute, 75Hz acquisition.
An electrophysiology system capable of providing multifocal stimulation, synchronising with the acquisition of data from a large number of electrodes and performing cross-correlation has been created. This is a powerful system which permits the interrogation of the dipoles evoked within the complex geometry of the visual cortex from a very large number of orientations, which will improve detection ability.
The system has been used to compare the performance of 16 monopolar recording channels in detecting responses to stimulation throughout the visual field. A selection of four electrodes which maximise the available information throughout the visual field has been made. It is shown that a several combinations of four electrodes provide good responses throughout the visual field, but that it is important to have them distributed on either hemisphere and above and below Oz.
A series of investigations have indicated methods of maximising the available information in mfVECP recordings and progress the technique towards becoming a robust clinical tool. A powerful multichannel multifocal electrophysiology system has been created, with the ability to simultaneously acquire data from a very large number of bipolar recording channels and thereby detect many small dipole responses to stimulation of many small areas of the visual field. This will be an invaluable tool in future investigations.
Performance has been shown to improve when the presence or absence of a waveform is determined by a novel SNR metric, when data is filtered post-acquisition through a 3-20Hz bandpass after cross-correlation and when a CRT is used to deliver the stimulus. The field of view of stimulation can usefully be extended to a radius of 30° when a 60-region dartboard pattern is employed. Performance can be enhanced at the same time as acquisition time is reduced by 25%, by the use of a 25Hz rate of stimulation instead of the frequently employed rate of 75Hz.

Item Type: Thesis (PhD)
Qualification Level: Doctoral
Keywords: Multifocal Visual Evoked Cortical Potential, mfVECP, neuroophthalmology, visual field, electrophysiology.
Subjects: R Medicine > RE Ophthalmology
Colleges/Schools: College of Medical Veterinary and Life Sciences > School of Cardiovascular & Metabolic Health
Supervisor's Name: Keating, Dr David
Date of Award: 2008
Depositing User: Mrs Jennifer A Macfarlane
Unique ID: glathesis:2008-317
Copyright: Copyright of this thesis is held by the author.
Date Deposited: 19 Aug 2008
Last Modified: 10 Dec 2012 13:17
URI: https://theses.gla.ac.uk/id/eprint/317

Actions (login required)

View Item View Item

Downloads

Downloads per month over past year