Puthenvila, George John (2026) Incorporating nitinol shape memory alloy in class IV flextensional transducers for future ultrasonic surgical devices. PhD thesis, University of Glasgow.
Full text available as:|
PDF
Download (17MB) |
Abstract
This research investigates the design, fabrication, material characterisation, and performance of class IV flextensional ultrasonic transducers incorporating Nitinol (Ni–Ti), a shape memory alloy (SMA) capable of reversibly transforming between martensitic and austenitic phases under the influence of temperature and applied stress. Previous studies have demonstrated the potential of Nitinol for resonance frequency tuning in class V cymbal transducers and enhanced thermal stability in Langevin ultrasonic transducers. The aim of this work was to evaluate the suitability of Nitinol for class IV flextensional ultrasonic transducers and to determine how its thermomechanical behaviour influences transducer performance under varying electrical, thermal, and mechanical operating conditions.
Material characterisation was first performed on the Nitinol components used throughout the study. Differential scanning calorimetry (DSC) was employed to determine transformation temperatures for Nitinol end-caps produced from materials with austenite finish temperatures (AF) of 35 °C and 45 °C, together with Nitinol wire possessing an AF of 30 °C. Energydispersive X-ray spectroscopy (EDXS) was used to evaluate elemental composition, while dynamic mechanical analysis (DMA) and tensile testing were conducted to investigate temperature-dependent mechanical behaviour, damping characteristics, and static mechanical performance. The results demonstrated that the thermomechanical response of Nitinol is influenced by both transformation behaviour and material processing history, highlighting the importance of material characterisation when integrating shape memory alloys into ultrasonic devices.
Class IV flextensional transducers incorporating Nitinol, aluminium (1050A), titanium (Ti-6Al-4V), and stainless steel (grade 316) end-caps were designed, fabricated, and characterised. Finite element analysis (FEA), electrical impedance analysis (EIA), resonance frequency tracking, displacement amplitude measurements, mechanical quality factor evaluation, and temperature-dependent testing were performed to assess transducer performance. While the conventional metallic transducers generally exhibited reductions in resonance frequency and mechanical quality factor with increasing temperature, the Nitinol transducer demonstrated greater resonance stability and reduced sensitivity to thermal variations. Although distinctive temperature-dependent behaviour was observed, the thermally induced resonance frequency tuning previously reported for Nitinol-based cymbal transducers was not observed within the flextensional transducer configurations investigated in this study.
To further investigate the influence of shape memory alloy integration, Nitinol wires were incorporated into the flextensional transducers and evaluated using impedance-matching circuitry. The wire-integrated transducers were characterised under varying voltage and temperature conditions to assess changes in resonance frequency, electrical impedance, mechanical quality factor, and power transfer characteristics. The results indicated that wire attachment influenced the electromechanical response of the transducers, while stable operation was maintained.
Ultrasonic surgical cutting was then conducted using the wire-integrated transducers. Cutting experiments were performed on bone-mimicking Sawbones® material while simultaneously monitoring resonance frequency, electrical impedance, phase, and load power. Among the investigated configurations, the Nitinol transducer with wire achieved the greatest measured cutting depth of 627.9 µm and exhibited one of the most stable electromechanical responses during prolonged operation. The titanium and stainless steel transducers with wires produced comparable cutting depths, while the aluminium transducer with wire exhibited the lowest cutting performance. Although the cutting depths achieved were low, the results demonstrated the feasibility of wire-integrated flextensional transducers for ultrasonic cutting applications and provided insight into the relationship between resonance stability and material removal.
Overall, this work demonstrates that Nitinol can be successfully integrated into class IV flextensional ultrasonic transducers and can influence their thermal response, dynamic behaviour, and cutting performance. While controllable frequency tuning was not demonstrated using the investigated configurations, the enhanced resonance stability exhibited by the Nitinol transducers suggests that shape memory alloys remain promising materials for future adaptive ultrasonic devices. The findings provide a foundation for the continued development of smart-material-integrated ultrasonic transducers for surgical and industrial applications.
| Item Type: | Thesis (PhD) |
|---|---|
| Qualification Level: | Doctoral |
| Subjects: | T Technology > TJ Mechanical engineering and machinery |
| Colleges/Schools: | College of Science and Engineering > School of Engineering |
| Supervisor's Name: | Lucas, Professor Margaret and Feeney, Professor Andrew |
| Date of Award: | 2026 |
| Depositing User: | Theses Team |
| Unique ID: | glathesis:2026-86384 |
| Copyright: | Copyright of this thesis is held by the author. |
| Date Deposited: | 20 Nov 2026 15:51 |
| Last Modified: | 22 Nov 2026 09:24 |
| Thesis DOI: | 10.5525/gla.thesis.86384 |
| URI: | https://theses.gla.ac.uk/id/eprint/86384 |